Real-Time, On-Line Solvent Vapor Measurements
Using NIR Tunable Diode Laser Absorption Spectroscopy

Advantages
- ✔ Gas species specific
- ✔ Continuous, real-time, high detection sensitivity
- ✔ Absolute solvent concentration measurements
- ✔ Integrated lot measurement
- ✔ Less susceptible to sampling errors
- ✔ Sampling uses existing system pump and pressure differential
- ✔ Remote operator control
- ✔ Limits worker exposure to active pharmaceutical ingredients
- ✔ Reliable: solid-state diode laser and fiber optic transport

System Configuration for Solvent Drying Applications

Custom-Designed Industrial, Rack Mounted, PSI Solvent Vapor Monitor
TDLAS Principles
-
Gas molecules absorb light at unique and discrete wavelengths
-
Laser wavelength repeatedly scans across absorption lines
- produces amplitude modulation in received power
-
The modulated signal is processed and the output translates into:
- molecules per cm3
- parts per million, PPM
- milligrams per liter
Typical Detection Limits*
(ppm at 1 atm total pressure, using a 1 meter optical path)
| Gas |
|
Gas |
|
| CH3OH |
20 |
H2O |
1 |
| CH2Cl2 |
10 |
CH3CN |
10 |
| CH3CHOHCH3 |
20 |
CH3CH2OH |
20 |
| NH3 |
5 |
HCl |
0.2 |
| O2 |
50 |
CO2 |
40 |
| H2S |
20 |
HCN |
1 |
| HF |
0.2 |
CO |
40 |
*Not a complete listing of possible detection gases, please inquire about your gas measurement needs.
|